Acoustic Noise and MRI Procedures

Various types of acoustic noise are produced during the operation of an MR system. Problems associated with acoustic noise for patients and healthcare professionals include annoyance, verbal communication difficulties, heightened anxiety, temporary hearing loss and, in extreme cases, the potential for permanent hearing impairment.

Acoustic noise may pose a particular problem to specific patient groups. For example, patients with psychiatric disorders may become confused or suffer from increased anxiety because of exposure to loud noise. Sedated patients may experience discomfort in association with high noise levels.

In addition, neonates may have adverse reactions to acoustic noise. Reeves MJ, et al. (2010) conducted a study to address this issue. The findings suggested that exposure of the fetus to 1.5-T MR imaging during the second and third trimesters of pregnancy is not associated with an increased risk of substantial neonatal hearing impairment or cochlear injury.

HEARING AND THE IMPACT OF ACOUSTIC NOISE

The human ear is a highly sensitive wide-band receiver, with the typical frequency range for normal hearing being between 20-Hz to 20,000-Hz. The ear does not tend to judge sound powers in absolute terms, but assesses how much greater one power is than another. The logarithmic decibel scale, dB, is used when referring to sound power.

Noise is defined in terms of frequency spectrum (in Hz), intensity (in dB), and time duration. Noise can be steady-state, intermittent, impulsive, or explosive. Transient hearing loss may occur following exposure to loud noise, resulting in a temporary threshold shift (i.e. a shift in the audible threshold).

With regard to acoustic noise associated with MR imaging, Brummett, et al. (1988) reported temporary shifts in hearing thresholds in 43% of the patients scanned without ear protection or with improperly fitted earplugs. Recovery from the effects of noise occurs in a relatively short period of time. However, if the noise insult is particularly severe, full recovery can take up to several weeks. If the noise is sufficiently injurious, a permanent threshold shift at specific frequencies may occur.

MRI-RELATED ACOUSTIC NOISE

The gradient magnetic field is the main source of acoustic noise associated with an MR procedure. This noise occurs during the rapid alterations of currents within the gradient coils. These currents, in the presence of the strong static magnetic field of the MR system, produce significant (Lorentz) forces that act upon the gradient coils. Acoustic noise, manifested as loud tapping, knocking, chirping, squeaking sounds, or other sounds is produced when the forces cause motion or vibration of the gradient coils as they impact against their mountings which, in turn, flex and vibrate.

Alteration of the gradient output (rise time or amplitude) by modifying MR imaging parameters causes the acoustic noise to vary. Noise tends to be enhanced by decreases in section thickness, field of view, repetition time, and echo time. In addition to dependence on imaging parameters, acoustic noise is dependent on the MR system hardware, construction, and the surrounding environment. Furthermore, noise characteristics have a spatial dependence. For example, noise levels can vary by as much as 10 dB as a function of patient position within the bore of the MR system. The presence and size of the patient may also affect the level of acoustic noise.

CHARACTERISTICS OF MR SYSTEM-RELATED ACOUSTIC NOISE

Gradient magnetic field-induced noise levels have been measured during a variety of pulse sequences for MR systems with static magnetic field strengths ranging from 0.35 to 4-Tesla. For example, Hurwitz, et al. (1989) reported that the MR imaging-related sound levels varied from 82 to 93-dB on the A-weighted scale and from 84- to 103-dB on the linear scale.

Later studies performed using a variety of MR parameters including “worst-case” pulse sequences that applied multiple gradients simultaneously (e.g., three-dimensional, fast gradient echo techniques) reported that these are among the loudest sequences, with acoustic noise levels that ranged from 103- to 113-dB (peak) on the A-weighted scale.

Additional studies measured acoustic noise generated by echo planar imaging (EPI) and fast spin echo sequences. Echo planar sequences tend to have extremely fast gradient switching times and high gradient amplitudes. Shellock, et al. (1998) reported high levels of noise ranging from 114- to 115-dBA on two different 1.5-Tesla MR systems tested during EPI sequences with parameters selected to represent “worst-case” protocols. At 3-Tesla, Hattori, et al. (2007) recorded sound levels that ranged from 126- to 131-dB on a linear scale, recommending the use of both earplugs and headphones for ear protection relative to the use of 3-Tesla MR systems when certain pulse sequences are used.

ACOUSTIC NOISE AND PERMISSIBLE LIMITS

In general, acoustic noise levels recorded in the MR environment have been below the maximum limits permitted by the Occupational Safety and Health Administration of the United States, especially when one considers that the duration of exposure is an important factor that determines the effect of noise on hearing.

The U.S. Food and Drug Administration released guidelines for acoustic noise levels that should not be exceeded in association with the operation of MR systems, as follows:

Sound Pressure Level - Peak unweighted sound pressure level greater than 140-dB. A-weighted root mean square (rms) sound pressure level greater than 99-dBA with hearing protection in place.

While the acoustic noise levels recommended for patients undergoing MR procedures on an infrequent and short-term basis may appear to be somewhat conservative, they are deemed appropriate when one considers that individuals with underlying health conditions may have problems with noise at certain levels or frequencies. Acoustic noise produced during MR procedures represents a potential risk to such patients. As previously mentioned, the possibility exists that substantial gradient magnetic field-induced noise may produce hearing problems in patients who are susceptible to the damaging effects of loud noises.

The exposure of staff and other healthcare workers in the MR environment is also a concern (e.g., those involved in interventional MR procedures or who remain in the room for patient management reasons). Accordingly, if loud noises exist in the MR environment, staff members should routinely wear hearing protection if they remain in the room during the operation of the scanner. In the United Kingdom, guidelines issued by the Department of Health recommend hearing protection be worn by staff exposed to an average of 85-dB over an eight hour day.

ACOUSTIC NOISE CONTROL TECHNIQUES

Passive noise control. The simplest and least expensive means of preventing problems associated with acoustic noise during MR procedures is routinely use disposable earplugs or headphones. Earplugs, when properly used, can abate noise by 10- to 30-dB, which is usually an adequate amount of sound attenuation for the MR environment. The use of earplugs typically provides a sufficient decrease in acoustic noise that, in turn, is capable of preventing hearing problems. Therefore, for MR systems that generate substantial acoustic noise, patients should be required to wear these protective devices.

Unfortunately, passive noise control methods suffer from a number of limitations. For example, these devices hamper verbal communication with patients during the operation of the MR system. Additionally, standard earplugs are often too large for the ear canal of adolescents and infants. Importantly, passive noise control devices provide non-uniform noise attenuation over the hearing range. While high frequencies may be well attenuated, attenuation is often poor at low frequencies. This is problematic because, for certain pulse sequences, the low frequency range is where the peak MR imaging-related acoustic noise is generated.

Active Noise Control (ANC). A significant reduction in the level of acoustic noise caused by MR procedures has been accomplished using active noise cancellation. Controlling noise from a particular source by introducing “anti-phase noise” to interfere destructively with the noise source is not a new idea with regard to MR systems. For example, in 1989, Goldman, et al. combined passive noise control and an active noise control system (i.e. an active system built into a headphone) to achieve an average noise reduction of approximately 14-dB.

Advances in digital signal processing technology allow efficient active noise control systems to be realized at a moderate cost. The anti-noise system involves a continuous feedback loop with continuous sampling of the sounds in the noise environment so that the gradient magnetic field-induced noise is attenuated. It is possible to attenuate the pseudo-periodic scanner noise while allowing the transmission of vocal communication or music to be maintained.

“Quiet” Pulse Sequences. Several investigators have described the development of “quiet” pulse sequences which substantially decrease acoustic noise and are acceptable for MR imaging and functional MRI examinations. These techniques are particularly interesting insofar as they may be utilized with echo planar imaging and other pulse sequences.

Silent Scan. Recently, Alibek, et al. (2014) described the use of “Silent Scan” technology which uses decreased gradient excitation levels to minimize acoustic noise associated with MR imaging. The acoustic noise levels were reduced significantly using a “Silenz” sequence compared with a conventional pulse sequence, according to the investigators. In another study, Costagli, et al. (2015) reported that acoustic noise was dramatically reduced using “Silent” T1-weighted brain imaging at 7-Tesla compared with conventional imaging. Thus, new techniques now exist that reduce acoustic noise and improve patient comfort.

OTHER SOURCES OF MR SYSTEM-RELATED ACOUSTIC NOISE

RF hearing. When the human head is subjected to pulsed radiofrequency (RF) radiation at certain frequencies, an audible sound perceived as a click, buzz, chirp, or knocking noise may be heard. This acoustic phenomenon is referred to as “RF hearing”, “RF sound” or “microwave hearing”.

Thermoelastic expansion is believed to be responsible for the production of RF hearing, whereby there is absorption of RF energy that produces a minute temperature elevation (i.e. approximately 1 x 10-6 degrees C) over a brief time period in the tissues of the head. Subsequently, a pressure wave is induced that is sensed by the hair cells of the cochlea via bone conduction. In this manner, a pulse of RF energy is transferred into an acoustic wave within the human head and sensed by the hearing organs.

With specific reference to the operation of MR scanners, RF hearing has been found to be associated with frequencies ranging from 2.4- to 170-MHz. The gradient magnetic field-induced acoustic noise that occurs during an MR procedure is significantly louder than the sounds associated with RF hearing. Therefore, noises produced by this RF auditory phenomenon are effectively masked and not perceived by patients. Currently, there is no evidence of any detrimental health effect related to the presence of RF hearing.

Noise From Subsidiary Systems. Patient comfort fans and cryogen reclamation systems associated with superconducting magnets of MR systems are the main sources of ambient acoustic noise found in the MR environment. Acoustic noise produced by these subsidiary systems is considerably less than that caused by gradient magnetic fields.

REFERENCES

Alibek S, et al. Acoustic noise reduction in MRI using Silent Scan: An initial experience. Diagn Interv Radiol 2014;20:360-3.

Bandettini PA, et al. Functional MRI of brain activation induced by scanner acoustic noise. Magn Reson Med 1998;39:410-416.

Bowtell RW, Mansfield PM. Quiet transverse gradient coils: Lorentz force balancing designs using geometric similitude. Magn Reson Med 1995;34:494-497.

Brummett RE, Talbot JM, Charuhas P. Potential hearing loss resulting from MR imaging. Radiology 1988;169:539-540.

Chaplain GBB. Anti-noise: The Essex breakthrough. Chart Mech Engin 1983;30:41-47.

Chen CK, Chiueh TD, Chen JH. Active cancellation system of acoustic noise in MR imaging. IEEE Trans Biomed Engin 1999;46:186-190.

Cho ZH, et al. Analysis of acoustic noise in MRI. Magnetic Resonance Imaging 1997;15:815-822.

Costagli M, et al. Assessment of Silent T1-weighted head imaging at 7 T. Eur Radiol 2015 Epub ahead of print.

Counter SA, Olofsson A, Grahn, Borg E. MRI acoustic noise: Sound pressure and frequency analysis. J Magn Reson Imag 1997;7:606-611.

Elder JA. Special senses. In: United States Environmental Protection Agency, Health Effects Research Laboratory. Biological Effects of Radio frequency Radiation. EPA-600/8-83-026F, Research Triangle Park, 1984, pp. 570-571.

Goldman AM, Gossman WE, Friedlander PC. Reduction of sound levels with anti-noise in MR imaging. Radiology 1989;173:549-550.

Hall DA, et al. Acoustic, psychophysical, and neuroimaging measurements of the effectiveness of active cancellation during auditory functional magnetic resonance imaging. J Acoust Soc Am. 2009;125:347-59.

Hattori Y, Fukatsu H, Ishigaki T. Measurement and evaluation of the acoustic noise of a 3 Tesla MR scanner. Nagoya J Med Sci 2007;69:23-28.

Hedeen RA, Edelstein WA. Characteristics and prediction of gradient acoustic noise in MR imagers. Magn Reson Med 1997;37:7-10.

Heismann B, et al. Sequence-based acoustic noise reduction of clinical MRI scans. Magn Reson Med 2015;73:1104-9

Hurwitz R, Lane SR, Bell RA, Brant-Zawadzki MN. Acoustic analysis of gradient-coil noise in MR imaging. Radiology 1989;173:545-548.

Ida M, et al. Quiet T1-weighted imaging using PETRA: Initial clinical evaluation in intracranial tumor patients. J Magn Reson Imaging 2015;41:447-53.

International Commission on Non-Ionizing Radiation Protection (ICNIRP) statement, medical magnetic resonance procedures: Protection of patients. Health Physics 2004;87:197-216.

Kaye EA, et al. Adapting MRI acoustic radiation force imaging for in vivo human brain focused ultrasound applications. Magn Reson Med 2013;69:724-33.

Lauer AM, et al. MRI acoustic noise can harm experimental and companion animals. J Magn Reson Imag 2012;36:743.

Lee J, et al. A real-time data acquisition and control of gradient coil noise for fMRI identification of hearing disorder in children with history of ear infection. Quant Imaging Med Surg 2013;3:28-42.

Leithner K, et al. Psychological reactions in women undergoing fetal magnetic resonance imaging. Obstet Gynecol 2008;111(2 Pt 1):396-402.

Lim EY, et al. 3 Tesla magnetic resonance imaging noise in standard head and neck sequence does not cause temporary threshold shift in high frequency. Eur Arch Otorhinolaryngol 2015;272;3109-13.

Lin CY, Chen JH. Real-time active noise control of magnetic resonance imaging acoustic noise. J Acoust Soc Am. 2012;132:1971.

Mansfield PM, Glover PM. Beaumont J Sound generation in gradient coil structures for MRI. Magn Reson Med 1998;39:539-550.

Mansfield PM, et al. Active acoustic screening: Design principles for quiet gradient coils in MRI. Meas Sci Technol 1994;5:1021-1025.

McJury MJ. Acoustic noise levels generated during high field MR imaging. Clinical Radiology 1995;50:331-334.

McJury M. Acoustic Noise and Magnetic Resonance Procedures. In: Magnetic Resonance Procedures: Health Effects and Safety. FG Shellock, Editor, CRC Press, Boca Raton, FL, 2001.

McJury M, Blug A, Joerger C, Condon B, Wyper D. Acoustic noise levels during magnetic resonance imaging scanning at 1.5 T. Brit J Radiol 1994;413-415.

McJury M, Shellock FG. Acoustic noise and MR procedures: A review. J Magn Reson Imag 2000;12:37-45.

McJury M, et al. The use of active noise control (ANC) to reduce acoustic noise generated during MRI scanning: Some initial results. Magnetic Resonance Imaging 1997;15:319-322.

Melnick W. Hearing loss from noise exposure. In: Harris CM, Editor. Handbook of Noise Control. New York: McGraw-Hill, 1979, pp. 2.

Miller LE, Keller AM. Regulation of occupational noise. In: Harris CM, Editor. Handbook of noise control. New York: McGraw-Hill, 1979, pp. 1-16.

Moelker A, et al. Verbal communication in MR environments: Effect of MR system acoustic noise on understanding. Radiology 2004;232:107-113.

Moelker A, Pattynama PM. Acoustic noise concerns in functional magnetic resonance imaging. Human Brain Mapping 2003;20:123-141.

More SR, et al. Acoustic noise characteristics of a 4 Tesla MRI scanner. J Magn Reson Imag 2006;23:388-397.

Ogurlu M, et al. Effect of headphones on sevoflurane requirement for MRI. Pediatr Radiol 2012;42:1432-6.

Peelle JE, et al. Evaluating an acoustically quiet EPI sequence for use in fMRI studies of speech and auditory processing. Neuroimage 2010;52:1410-9.

Philbin MK, Taber KH, Hayman LA. Preliminary report: Changes in vital signs of term newborns during MR. Am J Neuroradiol 1996;17:1033-6.

Pierre EY, et al. Parallel imaging-based reduction of acoustic noise for clinical magnetic resonance imaging. Invest Radiol 2014;49:620-6.

Quirk ME, Letendre AJ, Ciottone RA, Lingley JF. Anxiety in patients undergoing MR imaging. Radiology 1989;170:464-466.

Reeves MJ, et al. Neonatal cochlear function: Measurement after exposure to acoustic noise during in utero MR imaging. Radiology 2010;257:802-9.

Robinson DW. Characteristics of occupational noise-induced hearing loss. In: Effects of Noise on Hearing. Henderson D, Hamernik RP, Dosjanjh DS, Mills JH, Editors. Raven Press, New York, 1976, pp. 383-405.

Rondinoni C, et al. Effect of scanner acoustic background noise on strict resting-state fMRI. Braz J Med Biol Res 2013;46:359-67.

Roschmann P. Human auditory system response to pulsed radio frequency energy in RF coils for magnetic resonance at 2.4 to 170 MHz. Magn Reson Med 1991;21:197-215.

Ruckhäberle E, et al. In vivo intrauterine sound pressure and temperature measurements during magnetic resonance imaging (1.5 T) in pregnant ewes. Fetal Diagn Ther 2008;24:203-10.

Schmitter S, Bock M. Acoustic noise-optimized verse pulses. Magn Reson Med 2010;64:1446-52.

Segbers M, et al. Shaping and timing gradient pulses to reduce MRI acoustic noise. Magn Reson Med 2010;64:546-53.

Shellock FG, Kanal E. Policies, guidelines, and recommendations for MR imaging safety and patient management. J Magn Reson Imag 1991;1:97-101.

Shellock FG, Morisoli SM, Ziarati M. Measurement of acoustic noise during MR imaging: Evaluation of six “worst-case” pulse sequences. Radiology 1994;191:91-93.

Shellock FG, et al. Determination of gradient magnetic field-induced acoustic noise associated with the use of echo planar and three-dimensional fast spin echo techniques. J Magn Reson Imag 1998;8:1154-1157.

Strainer JC, et al. Functional MR of the primary auditory cortex: An analysis of pure tone activation and tone discrimination. Am J Roentgenol 1997;18: 601-610.

Strizek B, et al. Safety of MR imaging at 1.5 T in fetuses: A retrospective case-control study of birth weights and the effects of acoustic noise. Radiology 2015;275:530-7.

Tkach JA, et al. Characterization of acoustic noise in a neonatal intensive care unit MRI system. Pediatr Radiol 2014;44:1011-9

Tomasi DG, Ernst T. Echo planar imaging at 4 Tesla with minimal acoustic noise. J Magn Reson Imag 2003;128-130.

Ulmer JL, et al. Acoustic echo planar scanner noise and pure tone hearing thresholds: The effects of sequence repetition times and acoustic noise rates. J Comp Assist Tomogr 1998;22:480-486.

U.S. Department of Health and Human Services, Food and Drug Administration, Center for Devices and Radiological Health, Criteria for Significant Risk Investigations of Magnetic Resonance Diagnostic Devices, Guidance for Industry and Food and Drug Administration Staff, June 20, 2014.

Zapp J, et al. Sinusoidal echo-planar imaging with parallel acquisition technique for reduced acoustic noise in auditory fMRI. J Magn Reson Imag 2012;36:581-8.